This ion chamber came from Robert Millikan’s laboratory at the California Institute of Technology. Who used it and when it was used is unclear. Dr. Eugene “Bud” Cowan, who arranged the transfer to the ORAU collection, wrote that Victor Neher identified it as having been used by Millikan in the late 1920s - a note to that effect is written on the bottom of the wooden base. Nevertheless, when I sent Neher a photograph of this chamber, he did not recognize it. Despite this, the phrase “Victor Neher, 1950s” is written in pencil on the side of the chamber. |
|
|
What is certain is that the chamber is typical of those used at Millikan’s lab in the investigation of cosmic rays. In general, these chambers were attached to balloons that carried them to high altitudes (typically 70,000 – 80,000 ft.). This
particular chamber is a 10 inch diameter copper sphere with a thin
(1/8”) steel rod serving as the anode. When we opened it up for
inspection, we found that the chamber contained a few grams of silica gel
wrapped up in a fine copper screen. The
three-inch tube sticking out from the underside of the chamber was used to
evacuate and fill it with argon gas. The short projection at the top of
the sphere was there to hold the screw to which the line to the balloon
was attached. The system electronics were suspended by cords attached to
four connections located just below the chamber’s equator.
|
|
|
The electrical components attached to the wooden base below the chamber are a bit of a mystery. Nothing similar to them, as far as I have determined, is known to have been used at Cal Tech. As seen in the accompanying diagram, a metal extension of the chamber’s anode projects downward and ends in a small disk (ca. 0.3” in diameter). A flexible metal strip is located just below this disk. The pointed free end of the strip is bent upwards at right angles so that the tip points towards the circular disk. |
|
|
The
following explanation is my best guess as to how it was intended to
work - I doubt that it actually worked. It is possible that the normal
resting position of the flexible strip put its pointed metal tip in contact
with the circular metal disk connected to the chamber's anode. During use, a
constant charge was maintained on the strip. When the charge was first
applied, the anode was also charged and the similar charges on the anode and
metal strip would cause the latter to be deflected downward. This would
break the connection between the strip and anode and leave the latter
electrically isolated. As the chamber was exposed to cosmic rays, the
ionization of the gas inside the chamber would reduce the charge on the
anode thus reducing the repulsion between the anode and the strip.. The
metal strip would therefore gradually return to its resting position
eventually touching the disk. This would recharge the anode and the whole
process would be repeated. The frequency of this recharging of the anode was
used as the measure of the intensity of the cosmic radiation. A movie
camera would be used to record the movement of the strip.
The chambers normally used by Millikan and his coworkers operated in the essentially the same way. The major difference was that they employed a fine gold-plated quartz fiber that contacted the base of the anode instead of a crude metal strip that contacted a disk at the bottom of an extension of the anode. The use of cameras to record the movement of the fiber meant that the balloon had to be recovered after it came to earth, something that was not always possible. To solve this problem, Victor Neher designed a system in the early 1950s that used radio transmissions of the signals generated by the charging cycle. |
|
|
References Neher, H.V., An
Automatic Ionization Chamber, Rev.
Sci. Instr. 24 (2):99-102; 1953. Neher,
Peterson, and Stern, Cosmic Rays at High Altitudes and Latitudes,
Phys. Rev. 90 (4): 1953. |
|